Abstract
dc:descriptionPresent methods of monitoring, anesthetic breathing circuits for mechanical faults are flawed by inadequacies in the design of disconnect/pressure monitors (DPMs), devices which monitor breathing circuit pressure waveforms. Such inadequacies are investigated, and their cause is identified. A microprocessor based DPM, which applies pattern recognition techniques to the task of disconnect monitoring is proposed, and its design and implementation are presented. Technical and clinical trials show that the prototype DPM consistently detects almost all mechanical faults in the anesthetic breathing circuit. It is concluded that pattern recognition based DPMs are significantly more effective than their present day counterparts.
Degree
thesis:*- Name thesis:degree_name
- Master of Applied Science - MASc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Electrical and Computer Engineering
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1982
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Campbell, Michael A.
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/23742