Abstract
dc:descriptionThis thesis decribes the choice, the construction, the operation, and the testing of a ³He melting curve thermometer. The melting curve thermometer satisfies the need of this particular laboratory for a convenient standard thermometer in the temperature range of 6 mK to 600 mK. Over most of its operating range, the resolution of the temperature measured with the ³He melting curve thermometer is 0.03%, however, the accuracy is only 0.3%. The construction of a precision low temperature strain gauge is described. Also a ratio transformer capacitance bridge which has a resolution of 1 ppm is described.
Degree
thesis:*- Name thesis:degree_name
- Master of Science - MSc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1983
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Shinkoda, Ichiro
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/24773