Back to search

University of British Columbia

Improved noninvasive determination of blood pressure by oscillometry in the presence of motion artifacts

Abstract

dc:description

Blood pressure measurements provide information regarding the status of the cardiovascular system to aid in diagnosing pathological conditions, maintaining patient stability during surgery, and adapting tourniquet pressures. However, the performance of commercially available techniques of measuring blood pressure is often degraded during clinical measurements in the presence of motion artifacts. A prototype system based on oscillometry that uses algorithms which take advantage of a multi-bladder cuff has the potential to improve measurements of blood pressure in noisy environments by detecting noise and rapidly estimating blood pressure using only data uncorrupted by noise. This system and strategy, implemented in one version to track limb occlusion pressure (the minimum cuff pressure required to occlude underlying arteries), was evaluated along with a typical commercial, JLV available oscillometric blood pressure monitor to compare their ability to track limb occlusion pressure in the presence of varying noise conditions encountered during lab trials and clinical trials. Results showed that the prototype consistently estimated limb occlusion pressure more rapidly, more accurately, and more reliably than the oscillometric monitor in noise conditions typical of surgical procedures. The results also indicate that the prototype is feasible for incorporation into an adaptive tourniquet.

Degree

thesis:*
Name thesis:degree_name
Master of Applied Science - MASc
Level thesis:degree_level
master's
Discipline thesis:degree_discipline
Electrical and Computer Engineering
Grantor dc:publisher
University of British Columbia
Year dc:date
1986

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Bussani, Carlo Robert

Rights

dc:rights
Statement dc:rights
  • For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/2429/26218

Chain of custody

source
Harvested from
University of British Columbia
Base URL
circle.library.ubc.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Bussani, Carlo Robert. Improved noninvasive determination of blood pressure by oscillometry in the presence of motion artifacts. master's thesis, University of British Columbia, 1986. http://hdl.handle.net/2429/26218