Abstract
dc:descriptionA new technique has been devised to obtain time-resolved measurements of small spectral line shifts. Two linear neutral density wedges with opposite transmission gradients were placed over the top and bottom of a monochromator exit slit in such a way that a wavelength shift of a spectral line resulted in a change in light intensity transmitted through the wedges, which was monitored by two photomultipliers. To test this technique, an N II line, for which Griem has calculated the Stark shift as a function of electron density, was observed in the emission spectrum of a small theta-pinch using a monochromator of 10 Å/mm inverse dispersion. Space and time resolved line shift measurements were made and were associated with the axial velocity and the electron density fluctuations of the plasma. An uncertainty of 0.015 Å was estimated, which, for the 1.5 Å line used, corresponds to a shift to width ratio of 10⁻², believed to be lower than any previous shift measuring technique. Time of flight velocity measurements and conservation of mass considerations have confirmed the results as both reasonable and self-consistent.
Degree
thesis:*- Name thesis:degree_name
- Master of Science - MSc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1967
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Potter, Michael Urwin
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/36556