Back to search
University of British Columbia
The preparation and operation of lithium drift germanium detectors
Abstract
dc:descriptionLithium drifted germanium detectors have been prepared for use as high resolution gamma ray spectrometers. The fabrication procedure and the problems which can arise during preparation are discussed in detail. Using the techniques described, germanium detectors having the following characteristics were prepared. Active Volume 1.0 cm³ 0.5 cm³ 2.0 cm³ 1.7 cm³ Total Resolution at 661 keV 5.0 KeV 4.0 KeV 4.0 KeV 2.9 KeV
Degree
thesis:*- Name thesis:degree_name
- Master of Science - MSc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1966
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Thompson, Albert Charles
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/36563