The preparation and operation of lithium drift germanium detectors

Abstract

dc:description

Lithium drifted germanium detectors have been prepared for use as high resolution gamma ray spectrometers. The fabrication procedure and the problems which can arise during preparation are discussed in detail. Using the techniques described, germanium detectors having the following characteristics were prepared. Active Volume 1.0 cm³ 0.5 cm³ 2.0 cm³ 1.7 cm³ Total Resolution at 661 keV 5.0 KeV 4.0 KeV 4.0 KeV 2.9 KeV

Degree

thesis:*
Name thesis:degree_name
Master of Science - MSc
Level thesis:degree_level
master's
Discipline thesis:degree_discipline
Physics
Grantor dc:publisher
University of British Columbia
Year dc:date
1966

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Thompson, Albert Charles

Rights

dc:rights
Statement dc:rights
  • For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.