Calibration of orifice meters at low Reynolds numbers

Abstract

dc:description

Seventeen square-edged orifice plates were calibrated to study the effect of varying throat length, both in the absence and in the presence of geometrical similarity, on the coefficient of discharge for pipe Reynolds numbers from 20 to1 10,000. The experiments involved the use of hydraulically smooth pipe in test lines of 1 - inch, 1.5 - inch and 2 - inch nominal I.D., corner pressure taps and Beta ratios (orifice diameter/pipe diameter) of 0.2, 0.4 and 0.6. Six of the plates were built using design criteria recommended by the A.S.M.E. in "Fluid Meters" (1), and with Beta ratios specified at 0.4. Each test line was used with two plates, which were identical except for a variation in throat length within the range of the given design recommendations. A comparison of discharge coefficients as a function of pipe Reynolds number indicated the following: (1) Reynolds numbers 30 - 3,000. Consistent differences of 2-6% were observed in the calibration curves due to the variation in throat length allowed by "Fluid Meters".(1). (2) Reynolds number 3,000 - 10,000. Geometry appeared to be much less important, as all calibration curves tended to coincide. Two further plates, for the 1.5 - inch pipe only, were built using the A.S.M.E. "Power Test Code" (2) as a design reference. They were designed to have respectively the minimum and maximum throat length allowable under the code. The results, when plotted, indicated that in a pipe Reynolds number range of 40 - 1,500, plates built identically except for small differences in throat length, still gave distinguishably different calibration curves. The remaining nine plates were divided into three groups of three plates each, encompassing the three pipe diameters and three Beta ratios. All plates within a group were geometrically similar. The results, for pipe Reynolds numbers from 20 to 2,000, indicated that a total specification of orifice shape gave consistent calibration curves with no apparent absolute size effects. Thus, both "Fluid Meters" *and the "Power Test Code" recommended design criteria for standard orifice plates which do not completely take into account the effect of geometry on the coefficient of discharge at low Reynolds numbers. In particular, the tolerances allowed on throat length are too large.

Degree

thesis:*
Name thesis:degree_name
Master of Applied Science - MASc
Level thesis:degree_level
master's
Discipline thesis:degree_discipline
Chemical and Biological Engineering
Grantor dc:publisher
University of British Columbia
Year dc:date
1966

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Swain, Frederick Clifford

Rights

dc:rights
Statement dc:rights
  • For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/2429/37937

Chain of custody

source
Harvested from
University of British Columbia
Base URL
circle.library.ubc.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Swain, Frederick Clifford. Calibration of orifice meters at low Reynolds numbers. master's thesis, University of British Columbia, 1966. http://hdl.handle.net/2429/37937