University of British Columbia
Axial alignment in a ring-collection beta-ray spectrometer
Abstract
dc:descriptionA thin-lens beta-ray spectrometer using ring-focus collection was modified. These modifications consisted of; 1) a centering mechanism enabling the source-detector axis to be aligned with the magnetic axis; 2) an extension of the vacuum chamber placing the detector further from the magnet coils. The latter considerably decreased the magnetic shielding requirements for the detector. A misalignment of 0.25 mm. for parallel axes and of 0°09' for intersecting axes produced noticeably poorer performance. Using a gathering power of .70%, a resolving power of .94% was obtained for the 661.6 Kev. K-conversion peak of Cs ¹³⁷.
Degree
thesis:*- Name thesis:degree_name
- Master of Science - MSc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1960
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Earle, Eric Davis
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/40123