Abstract
dc:descriptionConstruction details of a 5" diameter 1.7 litre gas scintillation counter are given. The behaviour of the counter when containing He⁴ gas and a mixture of He⁴ and 10% Xe at various total pressures has been investigated. Two 5" photomultiplier tubes and a variety of different gas purifiers were tested in an attempt to obtain the best possible resolution. The pulse height was found to depend sensitively on the purification procedure adopted. A resolution of 10% for the Po²¹⁰ α-peak has been attained in two different chambers containing He⁴ plus 10% Xe. The voltage pulses rose in ≤300 nsec. Intense irradiation of a counter with 6 Mev x-rays caused only a slight increase in the maximum noise level already present from the photomultiplier tube. The behaviour of the counter as a fast neutron detector has been investigated using an uncollimated beam of 4.1 Mev neutrons. Pulses from recoiling protons and C¹² nuclei were observed as well as the He⁴ recoils.
Degree
thesis:*- Name thesis:degree_name
- Master of Science - MSc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1959
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Beer, George Atherley
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/40226