University of British Columbia
The construction and operating method of a new adiabatic calorimeter for measuring specific heats and heats of vaporization of liquids
Abstract
dc:descriptionIn the course of certain research work it became necessary to measure the specific heat of liquid cis and trans decahydrate with great accuracy; The ordinary Williams and Danielle adiabatic calorimeter was found to be impractical* Several years ago a calorimeter for measuring the specific heat of liquid hydrocarbons near room temperatures was developed at the U.S. Bureau of Standards, A study of this instrument proved that it would be the type desirable for the above measurement. Plans were obtained and altered to suit not only the measurements for cis and trans decalin but such hydrocarbons as are solid at room temperatures. During the past year such a calorimeter was constructed. It has also been so altered as to permit measurements of the latent heat of fusion to be made with 1 part in 1000.
Degree
thesis:*- Name thesis:degree_name
- Master of Applied Science - MASc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Chemical and Biological Engineering
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1948
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Selby, R. E.
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/41592