Back to search

University of Central Florida

Design of a bandpass oculometer

Abstract

dc:description.abstract

A passive oculometer is a device that measures the position of the eye by utilizing the corneo-retinal potential (CRP) generated by the eye. In this thesis, the Bandpass Oculometer replaces the joystick of a personal computer, allowing the viewer to use a computer simply by moving his eyes. Low frequency noise, present in all electronic circuits, produces an offset from the reference point, the center of the screen. In previous lowpass oculometry studies at University of Central Florida, this offset caused the cursor to drift toward the edge of the field of view within a period of 30 seconds, decreasing the ability to precisely control the cursor on the monitor. The Bandpass Oculometer is different from its predecessors in that bandpass filtering is utilized. By filtering all frequencies except where eye movement occurs, low frequency drift noise and higher frequency interference signals are eliminated. Consequently, the viewer can achieve a reference with no offset and no drift, resulting in smoother tracking and better control of the computer. An additional improvement to previous oculometers is the new computer interface. A diode shaping network is used to simulate a joystick resistance, and an optocoupler relays the output current to the computer. The optocoupler electrically isolates the viewer from the computer, thus making the system safer to use.

Degree

thesis:*
Grantor dc:publisher
University of Central Florida

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Brandauer, Karin
Contributors dc:contributor
  • Martin, Robert J.

Subjects

dc:subject × 9

Rights

dc:rights
Statement dc:rights
  • Public Domain
Language dc:language
English

Identifiers

dc:identifier.*
Identifier
DP0022053

Chain of custody

source
Harvested from
Central Florida
Base URL
stars.library.ucf.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation