UCL (University College London)
Glaucoma drainage devices: design, flow resistance and biocompatibility implications for intraocular pressure control
Abstract
dc:description.abstractThe overarching aim of this thesis is to develop a novel glaucoma drainage device that meets the ‘10-10-10’ goal for glaucoma treatment: specifically, a surgical intervention that can reduce the intraocular pressure to 10mmHg, can be implanted in 10 minutes and lasts for 10 years. The thesis encompasses elements from the research and design phase of device development to the preclinical and clinical testing of the device. The doctoral proposal is divided into four chapters. The first describes the role of minimally invasive surgery in the management of glaucoma; the second elucidates the physiochemical factors that impact the outflow facility of microtubes used in minimally invasive glaucoma surgery; the third presents a novel way to titrate aqueous outflow from a novel glaucoma microstent; finally, the fourth determines the bioburden, extractables and leachables of a novel glaucoma device, along with its implications for in vivo biocompatibility.
Degree
thesis:*- Level dc:type.qualificationlevel
- Doctoral
- Grantor dc:publisher.institution
- UCL (University College London)
- Year dc:date.issued
- 2021
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Henein, Christin
Rights
- Language dc:language
- en, eng
Identifiers
dc:identifier.*- Repository record dc:identifier.uri
- https://discovery.ucl.ac.uk/id/eprint/10120334/