Back to search

UCL (University College London)

Evaluation of digital x-ray detectors for medical imaging applications

Abstract

dc:description.abstract

Digital x-ray detectors are now the detector of choice in many X-ray examinations. They have been accepted into clinical practice over the past decade but there are still ongoing developments in the technology. Complementary metal oxide semiconductor (CMOS) active pixel sensors (APS) are a novel digital technology that offers advantages compared to some of the more established approaches (charge-coupled devices (CCD), thin film transistor arrays (TFT) and CMOS passive pixel sensors (PPS)). This thesis looks at the performance of these new sensors and attempts to identify their role in future medical imaging applications. Standard electro-optical and x-ray performance evaluations of two novel CMOS APS, namely the large Area Sensor (LAS) and Dexela CMOS x-ray detector, are presented. The evaluation was made in terms of the photon transfer curve (PTC), the modulation transfer function (MTF), the normalized noise power spectrum (NNPS) and the resultant detective quantum efficiency (DQE). Modifications were introduced to extend the standard methods to overcome technical limitations. The performance of these detectors was compared to three commercial systems (Remote RadEye HR (CMOS APS), Hamamatsu C9732DK (CMOS PPS) and Anrad SMAM (a-Se TFT)) at beam qualities (28 kV for mammography and 52 kV and 74 kV for general radiography) based on the IEC standards. Both the LAS and Dexela CMOS detectors demonstrate enhanced performance. The effect of the CMOS APS inherent nonlinearity on the x-ray performance was also evaluated. Finally, the measured performance parameters were used to simulate images for different mammographic imaging tasks in order to establish possible areas of application for the new sensors. Two software phantoms (one representing a 3-D breast and the other the CDMAM test tool) were used to simulate a range of mammographic conditions. The results show that both novel CMOS APS detectors offer high image quality compared to the commercial detector systems.

Degree

thesis:*
Level dc:type.qualificationlevel
Doctoral
Grantor dc:publisher.institution
UCL (University College London)
Year dc:date.issued
2011

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Konstantinidis, A.

Rights

Language dc:language
en, eng

Identifiers

dc:identifier.*
Repository record dc:identifier.uri
https://discovery.ucl.ac.uk/id/eprint/1322919/

Chain of custody

source
Harvested from
University College London
Base URL
discovery.ucl.ac.uk/cgi/oai2
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
related terms
citation

Konstantinidis, A.. Evaluation of digital x-ray detectors for medical imaging applications. Doctoral thesis, UCL (University College London), 2011. https://discovery.ucl.ac.uk/id/eprint/1322919/