Back to search

University of Minnesota

Progress Towards Development of a Desktop Oxygen Scanner for Assessing Macroencapsulated Islet Oxygenation

Abstract

dc:description.abstract

Islet macroencapsulation within an immunoisolation membrane is emerging as a promising new treatment for diabetes, yet providing adequate oxygenation to transplanted, encapsulated islets remains a major challenge because: 1) the islets require a specific level of oxygen (pO2) - both too little and too much oxygen impairs islet functionality and viability; 2) it is challenging to make in vivo oxygen measurements non-invasively. This thesis describes progress towards the development of a proto-type low-cost, portable, gradientless desktop MR oxygen scanner that can be used, in conjunction with an oxygen sensitive MR probe such as perfluoro-15-crown-5-ether (PFCE), to make non-invasive oxygen measurements in islet macroencapsulation devices in vivo. Novel methods are also described to accelerate oxygen measurement speed as well as to perform imaging and field mapping without using standard B0 gradients. The vision for this scanner, which is specifically designed for making pO2 measurements in macroencapsulation devices, is that one day it will be deployed in local pharmacies and supermarkets across the country, to meet the need that people with diabetes, who have received an islet macrencapsulation device implant, will have to get their oxygen measured and tuned for optimal transplant function.

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Wang, Paul

Subjects

dc:subject × 5

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Dc Identifier Citation
Wang, Paul. (2021). Progress Towards Development of a Desktop Oxygen Scanner for Assessing Macroencapsulated Islet Oxygenation. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/225125.

Chain of custody

source
Harvested from
University of Minnesota
Base URL
conservancy.umn.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Wang, Paul. Progress Towards Development of a Desktop Oxygen Scanner for Assessing Macroencapsulated Islet Oxygenation. 2021. https://hdl.handle.net/11299/225125