Development Of A Verified Distal End For 3D-Printed Transtibial Prosthetic Sockets

Abstract

dc:description.abstract

Background: An ideal lower limb prosthetic socket should be comfortable and durable while minimizing weight; optimizing socket designs for these goals can prove difficult due to the bespoke nature of definitive sockets. ISO 10328 is designed to verify the strength and durability of engineered prosthetic components but is commonly adapted for socket research. Sockets tested with ISO 10328 loading conditions typically fail at the distal end. Traditional socket fabrication methods involve downsides including long lead times, substantial labor, and variable quality brought on by manual fabrication. 3D-printing has been proposed as an alternative, but 3D-printed socket designs, most notably the distal end, are typically based on laminated sockets rather than taking advantage of the medium’s ability to generate complex geometry.Objectives: To design transtibial socket distal end geometry that can be used in 3D printed sockets across a variety of limb sizes and shapes, and to provide verification data from modified ISO 10328 testing that presents criteria for the merging radius between the socket body and distal end as well as evidence of the resulting load capacity. Methods: Using Topology Optimization, a distal end was modeled from simulated ISO 10328 conditions representing loads expected of a 125 kg user. Socket assemblies were designed using the distal end and tested with FEA to assess socket design criteria based on distal end build height and merging radius. Three example sockets were printed and tested under ISO 10328 ultimate strength criteria to the P6 level in Condition II for comparison to FEA results. Results: FEA testing suggested a minimum merging radius of 6 cm would produce a socket capable of withstanding P6 load levels. Build height should be minimized as much as possible without interfering with the merging radius. Experimental testing supports these conclusions, with sockets designed within these criteria withstanding the P6 ultimate strength maximum, and a socket violating the criteria failing at a lower load. All sockets failed in locations consistent with FEA modeling. Conclusions: Using Topology Optimization and FEA, 3D-printed prosthetic sockets that are both lightweight and durable were designed. The techniques and results presented offer a framework by which a distal end viable over a range of socket designs could be produced.

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Knoll, Ian Douglas

Subjects

dc:subject × 6

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Dc Identifier Citation
Knoll, Ian Douglas. (2023). Development Of A Verified Distal End For 3D-Printed Transtibial Prosthetic Sockets. Retrieved from the University Digital Conservancy, https://hdl.handle.net/11299/279183.

Chain of custody

source
Harvested from
University of Minnesota
Base URL
conservancy.umn.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Knoll, Ian Douglas. Development Of A Verified Distal End For 3D-Printed Transtibial Prosthetic Sockets. 2023. https://hdl.handle.net/11299/279183