Development and Implementation of a DSP Based Air Detector System to Prevent Embolism During Hemodialysis Therapy

Abstract

dc:description

This thesis describes the design of a DSP based air detector system to prevent air embolism during Hemodialysis, which is a treatment option for kidney failure disease. Hemodialysis consists of removing blood from the body, filtering and treating the blood to remove toxic substances such as wastes and fluids, reestablishing proper chemical levels in the blood and returning the processed blood to the body. The functions of hemodialysis are performed through the use of a dialyzer, which is also known as an artificial kidney. During hemodialysis small air bubbles may infiltrate the tubing used during the therapy and combine to form larger air bubbles that are harmful to the patient. If an air bubble is large enough and enters the patient's circulatory system, the blood flow can be blocked and the patient can die by embolism. Most of the hemodialysis instruments in use today are equipped with air detection systems, which are based on analog design and digital microcontroll ers. This thesis presents a design method based strictly on DSP technology. The Motorola DSP 56824EVM was considered suitable for this biomedical application since its performance parameters include high-speed, multi-signal control capability, reliability and stability. These performance parameters are considered to be the most important when designing biomedical instruments dealing with human beings' life and safety. The objective of this research was the development and implementation of a DSP algorithm for the detection and measurement of the sizes of air bubbles in a fluid. In addition the algorithm had to possess the capability, when appropriate, to initiate protective and awareness measures such as triggering a tube clamp as well as activating visual and audio alarms. The air detection was accomplished by means of a commercial air detector module, which was based on piezo ceramic and ultrasound sensing. The function of the tubing clamp was to stop the fluid flow in the tubing and prevent an air bubble from entering the patient's circulatory system. A secondary goal of this research was to exploit the capability of the DSP 56824EVM and demonstrate its suitability for biomedical applications.

Degree

thesis:*
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Electrical Engineering
Grantor
University of South Florida
Year dc:date
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Nguyen, Nhat
Contributors dc:contributor
  • Wilfrido Moreno, Ph.D.
  • James T. Leffew, Ph.D.
  • Paris Wiley, Ph.D.

Subjects

dc:subject × 7

Rights

Language dc:language
EN-US

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.usf.edu/etd/792

Chain of custody

source
Harvested from
University of South Florida
Base URL
digitalcommons.usf.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Nguyen, Nhat. Development and Implementation of a DSP Based Air Detector System to Prevent Embolism During Hemodialysis Therapy. Thesis thesis, University of South Florida, 2005. https://digitalcommons.usf.edu/etd/792