Abstract
dc:descriptionHemodynamic monitoring is extremely important in the accurate measurement of vital parameters. Current methods are highly invasive or noncontinuous, and require direct access to the patient’s skin. This study intends to explore the modulated magnetic signature of blood method (MMSB) to attain blood flow information. This method uses an applied magnetic field to magnetize the iron in the red blood cells and measures the disturbance to the field with a magnetic sensor [1]. Exploration will be done by experimentally studying in-vitro, as well as simulating in COMSOL the alteration of magnetic fields induced by the flow of a magnetic solution. It was found that the variation in magnetic field is due to a high magnetization of blood during slow flow and low magnetization during rapid flow. The understanding of this phenomenon can be used in order to create a portable, non-invasive, continuous, and accurate sensor to monitor the cardiovascular system.
Degree
thesis:*- Level thesis:degree_level
- Thesis
- Discipline thesis:degree_discipline
- Mechanical Engineering
- Grantor
- University of South Florida
- Year dc:date
- 2010
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Sinatra, Francy L.
- Contributors dc:contributor
-
- Rajiv Dubey, Ph.D.
- Shiv-Shankar Sundaram, Ph.D.
- Rasim Guldiken, Ph.D.
- Phillip J. Hipol, MSM&AT
Subjects
dc:subject × 8Rights
- Language dc:language
- EN-US
Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.usf.edu/etd/3518