Back to search

Hosted by Utah State University Libraries

Dew Point Hygrometer With Constant Resistance Humidity Transducer

Abstract

dc:description.abstract

The hygroscopic and electrical resistance characteristics of lithium chloride are investigated. Then an evaluation is made of the lithium chloride characteristics that would be suitable in development of a dew point hygrometer with measurement accuracy of ± 0.3 C. The development of a 12 volt battery-powered hygrometer is presented with circuit details and performance characteristics. Lithium chloride phase transition hygrometers tend to oscillate under certain conditions. A damping control is presented as part of the circuit details and recordings of the response characteristics and effectiveness of damping are presented showing effective control of the oscillations with a wide range of humidity transients.

Degree

thesis:*
Name thesis:degree_name
Master of Science (MS)
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Plants, Soils, and Climate
Grantor dc:publisher
Hosted by Utah State University Libraries

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Campbell, Curtis B.
Contributors dc:contributor
  • Gaylen L. Ashcroft

Subjects

dc:subject × 3

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Utah State University
Base URL
digitalcommons.usu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Campbell, Curtis B.. Dew Point Hygrometer With Constant Resistance Humidity Transducer. Thesis thesis, Hosted by Utah State University Libraries, https://digitalcommons.usu.edu/etd/2957