Abstract
dc:description.abstractThe Technicon AutoAnalyzer II was evaluated for automatic quantitation of non-casein protein (NCP) in milk. The Lowry method for colorimetric measurement of proteins was adapted to the automated method and found to be accurate in the determination of non-casein protein in milk. The automated Lowry method obeyed Beer's law. Casein content was calculated by difference between total protein determined by infrared milk analysis (IRMA) and non-casein protein. In the analysis of twenty-five replicate milk samples, the standard deviation was 0.32 and the coefficient of variability was 0.90. The casein content as a percentage of total protein in the milk of cows in the Utah State University dairy herd was found to vary from 41.0 to 81.3%. The automated or continuous flow method was affected by as little as 0.05% (w/w) mercuric chloride, a chemical preservative used in milk, whereas potassium dichromate and formaldehyde had insignificant effects upon the assay.
Degree
thesis:*- Name thesis:degree_name
- Master of Science (MS)
- Level thesis:degree_level
- Thesis
- Discipline thesis:degree_discipline
- Nutrition, Dietetics, and Food Sciences
- Grantor dc:publisher
- Hosted by Utah State University Libraries
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- White, Robert Steven
- Contributors dc:contributor
-
- Gary H. Richardson
- C. A. Ernstrom
- J. C. Batty