A droplet-based microfluidic impedance flow cytometer for detection and quantification of microplastics in water

Abstract

dc:description.abstract

The purpose of this thesis is to design and test a new microfluidic device based on impedance cytometry principles for detecting and quantifying microplastics in droplets. To measure the size and concentration of microplastics, we created a flow-focusing droplet generator with coplanar electrodes. Polystyrene microbeads of four distinct sizes and three different concentrations were employed. The impedance measurements were taken with a lock-in amplifier at three distinct frequencies: 4.4 MHz, 11 MHz, and 22.5 MHz. ANOVA and k-NN classification were used to examine the findings. The signal phase at low and medium frequencies was proven to best differentiate between different sizes and concentrations. The functional groups (for example, carboxyl groups) on the surface of microplastics can significantly interfere with the signal phase. The ANOVA and KNN classification findings demonstrated that the microfluidic device is highly sensitive to the concentration of non-carboxylated polystyrene beads and can classify concentration levels with 95.0% accuracy. According to a proposed equivalent circuit, the presence of microplastics in droplets can significantly influence the properties of the double-layer capacitance produced at the interface of the droplet and the electrode, making lower frequencies better for discriminating between different concentrations. The prominence of parasitic capacitance at high frequencies makes measurements indeterminate. Finally, it was demonstrated that the droplets' resistive behaviour does not alter as much as the double-layer capacitance behaviour, making medium frequencies less discriminative.

Degree

thesis:*
Level thesis:degree_level
Master of Applied Science M.A.Sc.
Department thesis:degree_department
Department of Mechanical Engineering
Year dc:date.issued
2022

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Aghel, Mohammadreza
Advisor dc:contributor.supervisor
  • Hoorfar, Mina

Subjects

dc:subject × 4

Rights

dc:rights
Statement dc:rights
  • Available to the World Wide Web
Language dc:language.iso
en, English

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1828/14590

Chain of custody

source
Harvested from
University of Victoria (Canada)
Base URL
dspace.library.uvic.ca/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Aghel, Mohammadreza. A droplet-based microfluidic impedance flow cytometer for detection and quantification of microplastics in water. Master of Applied Science M.A.Sc. thesis, 2022. http://hdl.handle.net/1828/14590