Reducing Uncertainty in Head and Neck Radiotherapy with Plastic Robotics

Abstract

dc:description.abstract

One of the greatest challenges in achieving accurate positioning in head and neck radiotherapy is that the anatomy at and above the cervical spine does not act as a single, mechanically rigid body. Current immobilization techniques contain residual uncertainties that are especially present in the lower neck that cannot be reduced by setting up to any single landmark. The work presented describes the development of a radiotherapy friendly mostly-plastic 6D robotic platform for positioning independent landmarks, (i.e., allowing remote, independent positioning of the skull relative to landmarks in the thorax), including analysis of kinematics, stress, radiographic compatibility, trajectory planning, physical construction, and phantom measurements of correction accuracy. No major component of the system within the field of imaging or treatment had a measured attenuation value greater than 250 HU, showing compatibility with x-ray-based imaging techniques. Relative to arbitrary overall setup errors of the head (min = 1.1 mm, max = 5.2 mm vector error) the robotic platform corrected the position down to a residual overall error of 0.75 mm +/- 0.33 mm over 15 cases as measured with optical tracking. This device shows the potential for providing reductions to dose margins in head and neck therapy cases, while also reducing setup time and effort.

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy
Level thesis:degree_level
Dissertation
Discipline thesis:degree_discipline
Medical Physics

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Ostyn, Mark R
Contributors dc:contributor
  • Siyong Kim
  • Woon-Hong Yeo
  • Laura Padilla
  • Taeho Kim

Subjects

dc:subject × 6

Rights

dc:rights
Statement dc:rights
  • © Mark Ostyn

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Virginia Commonwealth University
Base URL
scholarscompass.vcu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Ostyn, Mark R. Reducing Uncertainty in Head and Neck Radiotherapy with Plastic Robotics. Dissertation thesis, https://doi.org/10.25772/2XET-4B80