Back to search

Virginia Commonwealth University

Dosimetric Validation of an Electron Monte Carlo (eMC) Algorithm for Small Cutouts and Extended SSD’s: Assessing eMC in Small Field Dosimetry

Abstract

dc:description.abstract

This thesis validates the accuracy and limitations of Electron Monte Carlo (eMC) simulations in Eclipse treatment planning systems for small-field electron dosimetry and extended source-to-surface distances (SSDs). Systematic measurements were performed using a microDiamond detector across various electron energies and field sizes. The study examines the discrepancies between measured and calculated (eMC) dose distributions, validating challenges due to lateral scatter disequilibrium and extended air gaps. Our findings provide systematic data for eMC dose calculations and demonstrate that eMC performs within clinical tolerances, with the ultimate goal of enhancing the precision and safety of radiation treatments.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Radiation Oncology

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Basavaaraju, Anil
Contributors dc:contributor
  • Dr. Siyong Kim
  • Dr. Lulin Yuan
  • Dr. William Song

Subjects

dc:subject × 5

Rights

dc:rights
Statement dc:rights
  • © Anil Basavaraju

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Virginia Commonwealth University
Base URL
scholarscompass.vcu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Basavaaraju, Anil. Dosimetric Validation of an Electron Monte Carlo (eMC) Algorithm for Small Cutouts and Extended SSD’s: Assessing eMC in Small Field Dosimetry. Thesis thesis, https://doi.org/10.25772/8VM5-BC02