Virginia Commonwealth University
Dosimetric Validation of an Electron Monte Carlo (eMC) Algorithm for Small Cutouts and Extended SSD’s: Assessing eMC in Small Field Dosimetry
Abstract
dc:description.abstractThis thesis validates the accuracy and limitations of Electron Monte Carlo (eMC) simulations in Eclipse treatment planning systems for small-field electron dosimetry and extended source-to-surface distances (SSDs). Systematic measurements were performed using a microDiamond detector across various electron energies and field sizes. The study examines the discrepancies between measured and calculated (eMC) dose distributions, validating challenges due to lateral scatter disequilibrium and extended air gaps. Our findings provide systematic data for eMC dose calculations and demonstrate that eMC performs within clinical tolerances, with the ultimate goal of enhancing the precision and safety of radiation treatments.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- Thesis
- Discipline thesis:degree_discipline
- Radiation Oncology
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Basavaaraju, Anil
- Contributors dc:contributor
-
- Dr. Siyong Kim
- Dr. Lulin Yuan
- Dr. William Song
Subjects
dc:subject × 5Rights
dc:rights- Statement dc:rights
-
- © Anil Basavaraju
Identifiers
dc:identifier.*- Identifier
- https://scholarscompass.vcu.edu/etd/8006