Reaction monitoring using SABRE-hyperpolarised benchtop (1 T) NMR spectroscopy

Abstract

dc:description.abstract

This thesis develops novel methods for reaction monitoring using a low field (1 T, 43 MHz) benchtop NMR spectrometer combined with hyperpolarisation using the signal amplification by reversible exchange (SABRE) technique. Industry-friendly benchtop NMR spectrometers suffer from insensitivity, which is overcome here by hyperpolarisation. This work is focused on N-heterocyclic primary amines and amides as the target SABRE substrates. The optimisation of substrate concentration and polarisation transfer field was performed for individual compounds and their mixtures. Reaction monitoring studies were completed on an amide formation reaction, specifically conversion of 4-aminopyridine (4-AP) with benzoic anhydride to form N-(4-aminopyridyl)benzamide (4-PBA), and the reactivity of the SABRE catalyst itself. The transformation of the pre-catalyst [Ir(Cl)(COD)(IMes)] with the substrates 4-AP and 4-MP under hydrogen to form [Ir(H)2(IMes)(substrate)3]Cl was investigated. In addition, hydrogen isotope exchange (HIE) on the substrate in the presence of the SABRE complexes was explored in methanol-d4, where the source of the deuterium was the solvent. A new method was developed to monitor this reactivity using the lifetime of the SABRE-hyperpolarised 1H NMR response. For this purpose, a single-shot hyperpolarisation lifetime (hyperT1) measurement based on variable flip angles was developed that takes just five times the T1 time constant, typically tens of seconds, to complete with high SNR. This method was implemented with both high-field (9.4 T, 400 MHz) and low-field (1 T, 43 MHz) 1H NMR detection and comparable results were obtained using both instruments. In addition, the low sensitivity of the benchtop (1 T) NMR enabled the use of a protio solvent, which allowed for the differentiation between the effects of the catalyst formation and substrate deuteration. Calibration SABRE lifetime measurements fitted well to an analytical model of the kinetics of the SABRE process proposed in the literature. However, the quantification of the reaction monitoring data using this calibration revealed limitations of this simplified kinetic model.

Degree

thesis:*
Name dc:type.qualificationname
Ph.D
Level dc:type.qualificationlevel
doctoral
Grantor dc:publisher.institution
University of York
Year dc:date.issued
2019

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Semenova, Olga
Advisors dc:contributor.advisor
  • Duckett, Simon
  • Halse, Meghan

Rights

Language dc:language.dcterms_iso639_2
eng

Identifiers

dc:identifier.*
Identifier
uk.bl.ethos.792089

Chain of custody

source
Harvested from
White Rose University Consortium
Base URL
etheses.whiterose.ac.uk/cgi/oai2
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
related terms
citation

Semenova, Olga. Reaction monitoring using SABRE-hyperpolarised benchtop (1 T) NMR spectroscopy. doctoral thesis, University of York, 2019.