This thesis presents the design, implementation, and validation of an 8-channel bidirectional neurostimulator IC with a highly-linear high-dynamic-range ADC-direct architecture for simultaneous recording and stimulation. Each channel hosts a novel highly-linear high-dynamic-range recording architecture capable of amplification and quantization of brains neural signals in the presence of large differential-mode and common-mode stimulation artifacts, as well as a fully-programmable 8-bit current-mode electrical stimulator. The architecture enables the possibility of a patient-specific stimulation therapy required for the next generation of implantable closed-loop neuro-stimulators used for treatment of various neurological disorders. The proposed design adopts an ADC-direct architecture employing a dual-loop SAR-assisted continuous-time delta-sigma ADC architecture for differential-mode stimulation artifacts and offset removal. The presented channel achieves a high input impedance (1.8 G at 1 kHz), 400 mV linear input signal range, 94 dB dynamic range, and consumes 4.6 W with a signal bandwidth of 5 kHz.

Degree

thesis:*
Name thesis:degree_name
MASc - Master of Applied Science
Level thesis:degree_level
Master's
Discipline thesis:degree_discipline
Electrical and Computer Engineering
Year dc:date.issued
2021

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Moeinfard, Tania
Advisor dc:contributor.advisor
  • Kassiri, Hossein

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • Author owns copyright, except where explicitly noted. Please contact the author directly with licensing requests.
Language dc:language
en

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/10315/38160

Chain of custody

source
Harvested from
York University
Base URL
yorkspace.library.yorku.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Moeinfard, Tania. An 8-Channel Bidirectional Neurostimulator IC with a Highly-Linear High-Dynamic-Range ADC-Direct Architecture for Simultaneous Recording and Stimulation. Master's thesis, 2021. https://hdl.handle.net/10315/38160