• Refine Query
  • Source
  • Publication year
  • to
  • Language
  • 1
  • Tagged with
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • 1
  • About
  • The Global ETD Search service is a free service for researchers to find electronic theses and dissertations. This service is provided by the Networked Digital Library of Theses and Dissertations.
    Our metadata is collected from universities around the world. If you manage a university/consortium/country archive and want to be added, details can be found on the NDLTD website.
1

[en] METROLOGICAL ANALYSIS OF THE TRANSIENT BEHAVIOR OF A WET BULB AND DRY BULB PSYCHROMETER / [pt] ANÁLISE METROLÓGICA DO COMPORTAMENTO TRANSIENTE DE UM PSICRÔMETRO DO TIPO BULBO ÚMIDO E BULBO SECO

FRANKY EDUARDO REYES CAC 29 July 2005 (has links)
[pt] Este trabalho apresenta um estudo do comportamento transiente de um medidor de umidade tipo bulbo úmido e bulbo seco, assim como a determinação da umidade do ar em diferentes condições ambientais, que foram simuladas em laboratório. O psicrômetro do tipo bulbo úmido e bulbo seco é amplamente utilizado na indústria de ar condicionado, na monitoração da umidade em ambientes fechados, nas estações de meteorologia para a monitoração da umidade da atmosfera, e na prevenção das condições climáticas. Para o desenvolvimento do presente trabalho, foi construído um psicrômetro do tipo bulbo úmido e bulbo seco de ar aspirado, utilizando dois termômetros separados por uma placa de acrílico. Os termômetros utilizados foram tipo PT-100 e tipo Termopar. Na avaliação do conteúdo de umidade do ar, foi utilizada a medição de temperatura registrada por dois termômetros que compõem o psicrômetro, onde um deles registra a temperatura do ar que incide sobre o psicrômetro e serve como referência; e o segundo termômetro, recoberto com uma mecha molhada com água destilada, registra uma temperatura mais baixa devido ao fenômeno de resfriamento evaporativo, o qual é conseqüência da transferência de calor e massa entre a mecha molhada e o ar forçado a passar através dela. Na avaliação do comportamento transiente do psicrômetro, foram aplicados diferentes modelos de ajuste aos dados experimentais, procurando aquele que melhor representasse o comportamento através do tempo do termômetro bulbo úmido. Entre os modelos de ajuste estão: o linear, o exponencial, a soma de dois exponenciais, o racional, o polinomial de grau 6 e o exponencial linear (soma de exponencial com linha reta). A cada um dos modelos foi aplicado o método dos mínimos quadrados, e foi determinado por meio do critério de Akaike que o modelo exponencial linear é o melhor modelo de ajuste. As incertezas das constantes dos diferentes modelos de ajuste foram calculadas através do método das diferenças finitas. São apresentados os valores das constantes de tempo para diferentes valores de velocidade do ar e diferentes graus de umidade, tanto para os termômetros PT-100 quanto para o tipo Termopar. Finalmente, são apresentados os resultados da aplicação da regressão linear múltipla entre as constantes de tempo, a velocidade e a umidade do ar, tanto para o psicrômetro baseado nos termômetros tipo PT-100 quanto nos termômetros tipo Termopar. A presente pesquisa de mestrado contribui para o desenvolvimento da metrologia térmica no Brasil. / [en] This work presents a study of the transient behavior of a wet bulb and dry bulb psychrometer, as well as the determination of the humidity of air under different environment conditions. The wet bulb and dry bulb psychrometer are widely used in the air conditioning industry, for monitoring the humidity in closed environments, and in meteorology stations for the monitoring the atmosphere humidity and in the prevention of the climatic conditions. For the development of the present work, a wet bulb and dry bulb aspirated psychrometer was built, using two thermometers, which are physically separated by an acrylic plate. Two types of thermometers were used: PT-100 and Thermocouple. In the evaluation of the humidity content, the measurement of temperature was been registered by two thermometers (used as the psychrometer components), where one of them registers the temperature of the air that impacts the psychrometer and serves as reference; and the other one covered by a wick wet with distilled water, registers a lower temperature due to the phenomenon of evaporative cooling, which is a consequence of the mass transfer between the wet wick and the air that is forced to pass through it. In the evaluation of the transient behavior of the psychrometer different models of adjustment to the experimental data have been applied, seeking the best theoretical model that represented the behaviour through the time of the humid bulb thermometer. Among the adjustment models are: the linear one, the exponential one, the addition of two exponential ones, the rational one, polynomial of exponential degree 6 and the linear one (which adds the exponential curve and the straight line). For each of the models the method of the minimum square had been applied, and it was found by means of the of Aikaike´s criterion that the linear exponential model is the best one. The uncertainties of the constants of the different models of adjustment were calculated through the method of the finite differences. The values of the time constants of time for different values of air speed and different degrees of humidity for both thermometers PT-100 and Thermocouple type had been presented. Finally, the results of the application of the multiple linear regression are presented among the time constants, the speed, and the humidity of air for both psychrometer based on the thermometers PT- 100 type and the thermometers Thermocouple type. This MSc research contributes towards the development of the thermal metrology in Brazil.

Page generated in 0.0339 seconds