Spelling suggestions: "subject:"autoligature"" "subject:"ligaturant""
1 |
Étude comparative de la résistance au glissement de différents boîtiers orthodontiques selon l'inclinaison de deuxième ordreGratton, Frédéric January 2004 (has links)
Mémoire numérisé par la Direction des bibliothèques de l'Université de Montréal.
|
2 |
Une évaluation de la dimension de la lumière de boîtiers 0,018" provenant de quatre différents manufacturiers. Existe-t-il un standard dans l’industrie?Richard, Laurent A. 04 1900 (has links)
Introduction: The objective of this experimental research was to evaluate the slot’s vertical dimension and profile of four different 0.018″ self-ligating brackets and to identify the level of tolerance accepted by manufacturers during the fabrication process. It was then possible to calculate and compare the torque play of those brackets using the measured values and the nominal values.
Material and Methods: Twenty-five 0.018″ self-ligating brackets of upper left central incisors from the following manufacturers, Speed® (Strite Industries, Cambridge, Ontario, Canada), InOvationR® (GAC, Bohemia, NY, USA), CarriereLX® (Ortho Organizers, Carlsbad, CA, USA) and SmartClip® (3M Unitek, Monrovia, CA, USA), were evaluated using electron microscopy with 150X images. The height of each bracket was measured at every 100 microns of depth from the lingual wall at five different levels. A Student T test was then used to compare our results with the manufacturer’s stated value of 0.018″. To determine if there was a significant difference between the four manufacturers, analysis of variance (ANOVA) was performed at the significance level of p<0.05. The torque play was then calculated using geometrical formulas.
Results: On average, Speed brackets were oversized by 2.7%[MV 0.0185″ (SD:0.002)], InOvationR by 3.7% [MV 0.0187″ (SD:0.002)], CarriereLX by 3.2% [MV 0.0186″ (SD:0.002)] and SmartClipSL by 5.0% [MV 0.0189″ (SD:0.002)]. The height of all brackets was significantly higher than the nominal value of 0.018″ (p<0.001). The slot of SmartClip brackets was significantly larger than those of the other three manufacturers (p<0.001). None of the brackets studied had parallel gingival and occlusal walls; some were convergent and others divergent. These variations can induce a torque play up to 4.5 degrees with a 0.017″x0.025″ wire and 8.0 degrees with a 0.016″x0.022″ wire.
Conclusion: All studied brackets were oversized. None of the brackets studied had parallel gingival and occlusal walls and there was no standard between manufacturers for the geometry of their slots. These variations can cause a slight increase of the torque play between the wire and the bracket compared with the nominal value. / Introduction : Cette recherche a été une étude expérimentale conduite en laboratoire visant à évaluer la dimension verticale ainsi que la géométrie de la lumière de différents boîtiers auto-ligaturants populaires de grandeur 0,018″. Par la suite, il a été possible de calculer mathématiquement la perte de torque engendrée par ces variations.
Matériel et méthode : Vingt-cinq boîtiers auto-ligaturants d’incisive centrale supérieure gauche de quatre différents manufacturiers, Speed® (Strite Industries, Cambridge, Ontario, Canada), InOvationR® (GAC, Bohemia, NY, USA), CarriereLX® (Ortho Organizers, Carlsbad, CA, USA) et SmartClip SL® (3M Unitek, Monrovia, CA, USA), ont été évalués en utilisant un microscope à balayage électronique à un grossissement de 150X. La hauteur de la lumière a été mesurée à tous les 100 microns de profondeur à cinq différents niveaux, permettant ainsi d’établir sa hauteur moyenne et d’évaluer le parallélisme de ses parois. Le test T de Student a été utilisé pour déterminer si ces valeurs étaient statistiquement significatives comparativement à la valeur de référence de 0,018″. Un test de variance (ANOVA), avec un seuil de signification p<0,05, a été utilisé pour comparer les boîtiers des différents manufacturiers entre eux. La perte de torque a, par la suite, été calculée théoriquement à l’aide de formules géométriques en fonction d’un fil donné.
Résultats : En moyenne, la lumière des boîtiers Speed était surdimensionnée de 2,7% [VM : 0,0185″ (DS : 0,002)], InOvationR de 3,7% [VM 0,0187″ (DS : 0,002)], CarriereLX de 3,2% [VM 0,0186″ (DS : 0,002)] et SmartClip SL de 5,0% [VM 0,0189″ (DS : 0,002)]. La hauteur moyenne des quatre types de boîtiers a démontré une différence statistiquement significative par rapport à la valeur de référence de 0,018″ (p<0,001). Il existe également une différence statistiquement significative entre la hauteur moyenne des boîtiers SmartClip et celle des trois autres manufacturiers (p<0,001). Aucun des boîtiers étudiés n’avait les parois gingivales et occlusales parallèles, certaines étaient divergentes, d’autres convergentes. Ces variations peuvent engendrer cliniquement une perte de torque jusqu’à 4,5 degrés avec un fil 0,017″x0,025″ et jusqu’à 8,0 degrés avec un fil 0,016″x0,022″.
Conclusion : La lumière de tous les boîtiers étudiés était légèrement surdimensionnée. La géométrie de la lumière variait d’un manufacturier à l’autre, certains boîtiers étaient convergents et d’autres divergents. Ces variations causées par les méthodes de fabrication engendrent une perte de torque supplémentaire par rapport à la valeur nominale.
|
3 |
Une évaluation de la dimension de la lumière de boîtiers 0,018″ provenant de quatre différents manufacturiers. Existe-t-il un standard dans l’industrie?Richard, Laurent A. 04 1900 (has links)
Introduction: The objective of this experimental research was to evaluate the slot’s vertical dimension and profile of four different 0.018″ self-ligating brackets and to identify the level of tolerance accepted by manufacturers during the fabrication process. It was then possible to calculate and compare the torque play of those brackets using the measured values and the nominal values.
Material and Methods: Twenty-five 0.018″ self-ligating brackets of upper left central incisors from the following manufacturers, Speed® (Strite Industries, Cambridge, Ontario, Canada), InOvationR® (GAC, Bohemia, NY, USA), CarriereLX® (Ortho Organizers, Carlsbad, CA, USA) and SmartClip® (3M Unitek, Monrovia, CA, USA), were evaluated using electron microscopy with 150X images. The height of each bracket was measured at every 100 microns of depth from the lingual wall at five different levels. A Student T test was then used to compare our results with the manufacturer’s stated value of 0.018″. To determine if there was a significant difference between the four manufacturers, analysis of variance (ANOVA) was performed at the significance level of p<0.05. The torque play was then calculated using geometrical formulas.
Results: On average, Speed brackets were oversized by 2.7%[MV 0.0185″ (SD:0.002)], InOvationR by 3.7% [MV 0.0187″ (SD:0.002)], CarriereLX by 3.2% [MV 0.0186″ (SD:0.002)] and SmartClipSL by 5.0% [MV 0.0189″ (SD:0.002)]. The height of all brackets was significantly higher than the nominal value of 0.018″ (p<0.001). The slot of SmartClip brackets was significantly larger than those of the other three manufacturers (p<0.001). None of the brackets studied had parallel gingival and occlusal walls; some were convergent and others divergent. These variations can induce a torque play up to 4.5 degrees with a 0.017″x0.025″ wire and 8.0 degrees with a 0.016″x0.022″ wire.
Conclusion: All studied brackets were oversized. None of the brackets studied had parallel gingival and occlusal walls and there was no standard between manufacturers for the geometry of their slots. These variations can cause a slight increase of the torque play between the wire and the bracket compared with the nominal value. / Introduction : Cette recherche a été une étude expérimentale conduite en laboratoire visant à évaluer la dimension verticale ainsi que la géométrie de la lumière de différents boîtiers auto-ligaturants populaires de grandeur 0,018″. Par la suite, il a été possible de calculer mathématiquement la perte de torque engendrée par ces variations.
Matériel et méthode : Vingt-cinq boîtiers auto-ligaturants d’incisive centrale supérieure gauche de quatre différents manufacturiers, Speed® (Strite Industries, Cambridge, Ontario, Canada), InOvationR® (GAC, Bohemia, NY, USA), CarriereLX® (Ortho Organizers, Carlsbad, CA, USA) et SmartClip SL® (3M Unitek, Monrovia, CA, USA), ont été évalués en utilisant un microscope à balayage électronique à un grossissement de 150X. La hauteur de la lumière a été mesurée à tous les 100 microns de profondeur à cinq différents niveaux, permettant ainsi d’établir sa hauteur moyenne et d’évaluer le parallélisme de ses parois. Le test T de Student a été utilisé pour déterminer si ces valeurs étaient statistiquement significatives comparativement à la valeur de référence de 0,018″. Un test de variance (ANOVA), avec un seuil de signification p<0,05, a été utilisé pour comparer les boîtiers des différents manufacturiers entre eux. La perte de torque a, par la suite, été calculée théoriquement à l’aide de formules géométriques en fonction d’un fil donné.
Résultats : En moyenne, la lumière des boîtiers Speed était surdimensionnée de 2,7% [VM : 0,0185″ (DS : 0,002)], InOvationR de 3,7% [VM 0,0187″ (DS : 0,002)], CarriereLX de 3,2% [VM 0,0186″ (DS : 0,002)] et SmartClip SL de 5,0% [VM 0,0189″ (DS : 0,002)]. La hauteur moyenne des quatre types de boîtiers a démontré une différence statistiquement significative par rapport à la valeur de référence de 0,018″ (p<0,001). Il existe également une différence statistiquement significative entre la hauteur moyenne des boîtiers SmartClip et celle des trois autres manufacturiers (p<0,001). Aucun des boîtiers étudiés n’avait les parois gingivales et occlusales parallèles, certaines étaient divergentes, d’autres convergentes. Ces variations peuvent engendrer cliniquement une perte de torque jusqu’à 4,5 degrés avec un fil 0,017″x0,025″ et jusqu’à 8,0 degrés avec un fil 0,016″x0,022″.
Conclusion : La lumière de tous les boîtiers étudiés était légèrement surdimensionnée. La géométrie de la lumière variait d’un manufacturier à l’autre, certains boîtiers étaient convergents et d’autres divergents. Ces variations causées par les méthodes de fabrication engendrent une perte de torque supplémentaire par rapport à la valeur nominale.
|
Page generated in 0.0466 seconds