Spelling suggestions: "subject:"testing off optical device"" "subject:"testing oof optical device""
1 |
Development and Testing of a Second Generation Hand-held Optical ImagerGonzalez, Jean 22 March 2012 (has links)
Hand-held optical imagers are developed towards clinical breast cancer imaging. Herein, a Gen-2 hand-held optical imager has been developed with unique features: (i) image curved breast tissues with ~86% surface contact, and (ii) perform reflectance and transillumination imaging using the novel forked probe heads. Extensive phantom studies were performed using 1% Liposyn solution (background, ~ 300 ml and 1000 ml volumes) and 0.45 cc India Ink (absorption) targets, under different target:background contrast ratios and target depths. Two-dimensional surface images detected target(s) up to 2.5 cm deep via reflectance imaging, and up to 5 cm deep via transillumination imaging. Preliminary studies on gel-based breast phantoms (~700 ml) detected targets via reflectance and transillumination imaging. Preliminary in-vivo reflectance studies on normal and cancerous breast tissues also detected targets, although with artifacts. In future, the portable Gen-2 imager has potential for clinical breast imaging via reflectance and transillumination approach after extensive in-vivo studies.
|
Page generated in 0.1028 seconds