Otimização do processo de descontaminação no sistema isolador de Bio-Manguinhos

Submitted by Priscila Nascimento (pnascimento@icict.fiocruz.br) on 2012-11-30T12:31:09Z
No. of bitstreams: 1
beatriz-cyranka.pdf: 1720638 bytes, checksum: c436adb348584d5a5a8213fccdef6ea6 (MD5) / Made available in DSpace on 2012-11-30T12:31:09Z (GMT). No. of bitstreams: 1
beatriz-cyranka.pdf: 1720638 bytes, checksum: c436adb348584d5a5a8213fccdef6ea6 (MD5)
Previous issue date: 2011 / Fundação Oswaldo Cruz. Instituto de Tecnologia em Imunobiológicos. Rio de Janeiro, RJ, Brasil. / Novas tecnologias, como o sistema isolador para ensaios de esterilidade vieram a
contribuir com a incorporação de um melhor desempenho destes ensaios no
controle de qualidade das indústrias farmacêuticas.Este trabalho teve como objetivo
a verificação do processo de biodescontaminação dentro do sistema isolador de Bio-Manguinhos, utilizado como agente esterilizante o gás de peróxido de hidrogênio.
Foram utilizadas como biocarga para redução de contaminação microbiológica três
concentrações dos microrganismos, Candida albicansATCC 10231, Clostridium
sporogenesATCC 11437 eMicrococcus luteusATCC 9341, impregnadas em discos
de filtro de celulose. Os estudos de cinética de crescimento dos microrganismos
foram realizados para um melhor entendimento do seumetabolismo, bem como
aspectos gerais de crescimento que contribuíram para ressaltar que a Candida
albicans inicia sua fase exponencial de crescimento na segunda hora do cultivo e
finaliza esta etapa na sexta hora do cultivo, com produção máxima de células
viáveis, fato observado também no microrganismo Micrococccus luteus. Para o
cultivo doClostridium sporogenes o crescimento foi mais lento com uma curva de
crescimento com 60 horas de cultivo total. A produção de maior número de células
para o Clostridiumfoi alcançada na vigésima quarta hora de cultivo, assim como a
maior produção de esporos. Foi estabelecida, ao longo da curva de crescimento, a
relação entre densidade ótica e número de células viáveis, relação essa importante
para o estabelecimento das condições do estudo em relação à biocarga empregada
de cada microrganismo no momento do desafio no sistema isolador. A capacidade
de descontaminação avaliada dentro do sistema isolador com o biocida peróxido de
hidrogênio revelou o tempo de exposição ao gás de 10 minutos, como resultado final
satisfatório apresentando redução total da carga microbiana com destruição total das
células viáveis, assim como as formas esporuladas do Clostridium sporogenes.
Desta forma conclui-se que o peróxido de hidrogênioé um biocida de eficácia
comprovada, nas variáveis deste estudo e o processode descontaminação no
sistema isolador de Bio-Manguinhos é compatível comsua atividade finalística na
produção de insumos para a saúde. / New technologies such as isolator system for sterility tests came to help with the
incorporation of a better performance of these kinds of tests in quality control of
pharmaceutical companies. This study aimed to verify the decontamination process
within the system isolator of Bio-Manguinhos, usinghydrogen peroxide gas as a
sterilizing agent. Three concentrations of microorganisms were used as bioburden for
the microbiological contamination reduction, Candida albicansATCC 10231,
Clostridium sporogenesATCC 11437 and Micrococcus luteusATCC 9341,
impregnated in cellulose filter discs. Studies of growth kinetics of microorganisms
were carried out to a better understanding of theirmetabolism, as well as general
aspects of growth that contributed to emphasize that the Candida albicansbegins its
exponential growth phase in the second hour of cultivation and this step ends at the
sixth hour cultivation, with maximum yield of viable cells, this was also observed in
the Micrococcus luteusmicroorganism. For the Clostridium sporogenescultivation,
growth was slower with a 60 hours growth curve of total culture. The production of
more cells for Clostridiumwas achieved in the twenty-fourth hour of cultivation, as
well as the maximum spores production. It was established along the growth curve
the relationship between optical density and numberof viable cells, this relationship
was important to establish the conditions of the study related to the bioburden of
each microorganism used to challenge the isolator system. The decontamination
capacity evaluated within the isolator system with the hydrogen peroxide biocide
showed that the gas exposure time of 10 minutes wassatisfactory demonstrating
total reduction of the microbial load with total destruction of viable cells, as well as
the sporulated forms of Clostridium sporogenes. Thus it is concluded that hydrogen
peroxide is a proved effective biocide, in the variables of this study and
decontamination process in the Bio-Manguinhos insulator system is compatible with
its main activity in the production of health supplies.

Identiferoai:union.ndltd.org:IBICT/oai:www.arca.fiocruz.br:icict/5910
Date January 2011
CreatorsCyranka, Beatriz
ContributorsLemes, Elezer Monte Blanco, Rezende, Cleide A. Ferreira de, Santos, Ana Paula dos, Matos, Denise Cristina de Souza, Senna, José Procópio Moreno
PublisherInstituto de Tecnologia em Imunobiológicos.
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguagePortuguese
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/masterThesis
Sourcereponame:Repositório Institucional da FIOCRUZ, instname:Fundação Oswaldo Cruz, instacron:FIOCRUZ
Rightsinfo:eu-repo/semantics/openAccess

Page generated in 0.015 seconds