[en] ANALYSIS OF THE INFLUENCE OF VISCOSITY ON THE PETROLEUM FLOW RATE MEASUREMENT BY ULTRASONIC METERS / [pt] ANÁLISE DA INFLUÊNCIA DA VISCOSIDADE NA MEDIÇÃO DE VAZÃO DE PETRÓLEO POR MEDIDORES ULTRASSÔNICOS

[pt] No Brasil, a medição de petróleo é regulada pela ANP e pelo INMETRO. O
Regulamento Técnico de Medição de Petróleo e Gás Natural exige que os
medidores de vazão de petróleo sejam calibrados nas condições mais próximas
das encontradas no local de medição, tanto para o fluido (massa específica e
viscosidade) quanto operacionais (vazão, pressão e temperatura). Replicar todas
estas condições nos laboratórios de calibração nacionais é muito complicado e
dispendioso. Assim, neste trabalho foram avaliados os comportamentos de 7
medidores Altosonic V calibrados com 5 tipos de fluidos (oural, condensat, fuel,
heavy fuel e água). Os fatores de calibração dos medidores foram submetidos a
avaliações estatísticas e curvas de calibração foram ajustadas buscando
identificar possíveis tendências de comportamento. Após supor-se que vazões e
viscosidades teriam influência na aleatoriedade dos fatores, chegou-se à
conclusão que, utilizando as calibrações de fábrica (fingerprint) do medidor, que
cobrem toda a faixa de viscosidade e vazão de medição com a utilização do
número de Reynolds, os limites requeridos por norma de erros para medição
fiscal de são atendidos. Assim, sugere-se que as verificações periódicas do
Altosonic V sejam feitas com o fluido disponível no laboratório de calibração,
observando a coerência dos fatores obtidos nesta calibração e na calibração
inicial, buscando o atendimento aos limites legais da aplicação. / [en] In Brazil, the measurement of oil is regulated by ANP and INMETRO. The
Technical Regulation for Petroleum and Natural Gas requires that the oil flow
meters be calibrated in conditions close to those found at the measurement site,
both regarding the fluid (density and viscosity) and operational conditions (flow
rate, pressure and temperature). It is very difficult to replicate all these conditions
in national calibration facilities and expensive too. Thus, this study evaluated the
behavior of 7 Altosonic V meters calibrated with 5 types of fluids (oural,
condensat, fuel, heavy fuel and water). The calibration factors of the meters were
subjected to statistical evaluation and calibration curves were fitted looking for
possible behavior trends. Supposing that flow rates and viscosities would
influence randomly the factors, a conclusion was reached that, by using the
factory meter calibrations (fingerprint), that cover the entire measurement range
of viscosity and flow rates, by using Reynolds number, the required by norm
fiscal measurement error limits are met. Thus, it is suggested that periodic
checks of Altosonic V could be made with the fluid available in the calibration
facility, verifying the consistency of the factors obtained in this calibration and in
the fingerprint, aimed at complying with the legal limits of the application.