Protótipo de ponteira acoplada aos aparelhos fotopolimerizadores para colagem de bráquetes e acessórios ortodônticos

Submitted by Renata Lopes (renatasil82@gmail.com) on 2016-05-18T14:18:46Z
No. of bitstreams: 1
sergioluizmotajunior.pdf: 65970440 bytes, checksum: b08afa740679beb93b85550945b417d2 (MD5) / Approved for entry into archive by Adriana Oliveira (adriana.oliveira@ufjf.edu.br) on 2016-07-01T18:47:10Z (GMT) No. of bitstreams: 1
sergioluizmotajunior.pdf: 65970440 bytes, checksum: b08afa740679beb93b85550945b417d2 (MD5) / Made available in DSpace on 2016-07-01T18:47:10Z (GMT). No. of bitstreams: 1
sergioluizmotajunior.pdf: 65970440 bytes, checksum: b08afa740679beb93b85550945b417d2 (MD5)
Previous issue date: 2012-02-14 / CAPES - Coordenação de Aperfeiçoamento de Pessoal de Nível Superior / O objetivo do presente estudo foi o desenvolvimento de uma nova ponteira para ser
acoplada aos aparelhos fotopolimerizadores utilizados para colagem de bráquetes e
acessórios ortodônticos, e teste da efetividade da mesma em ensaio mecânico in
vitro. A ponteira é espelhada na superfície interna e baseia-se em conceitos físicos
de refração e reflexão de luz. Apresenta como principal vantagem, a redução do
tempo clínico necessário para o procedimento de colagem, reduzindo, dessa forma,
a possibilidade de contaminação durante o processo. Através do ensaio de
resistência mecânica ao cisalhamento e da determinação do Indice Remanescênte
de Adesivo (IRA), testou-se a ponteira em uma amostra composta por 120 corpos de
prova, sendo a mesma dividida em 2 grupos. No grupo 1 foi utilizado aparelho
fotopolimerizador de fonte de luz halógena e no grupo 2 foi utilizado fonte de LED.
Cada grupo foi subdividido. Nos subgrupos H1 e L1 utilizou-se a ponteira
convencional dos respectivos instrumentos. Nos subgrupos H2 e L2 a colagem foi
feita utilizando-se a ponteira desenvolvida. Os valores dos testes de cisalhamento e
IRA para os subgrupos foram comparados entre si. Os resultados mostraram que
não houve diferença estatisticamente significante, tanto para os ensaios de
resistência ao cisalhamento (p>0,05), quanto para o IRA (p>0,05) entre os
subgrupos. Conclui-se que 1 - os testes de ensaios mecânicos, assim como a
análise do IRA mostraram que a nova ponteira desenvolvida, cumpriu os requisitos
necessários à colagem dos acessórios ortodônticos; 2 – o tempo de colagem foi
reduzido pela metade, sendo necessária uma só incidência. / The porpose of the present study was the development of a new device to be
coupled to light-curing units for bonding orthodontic brackets and accessories and
test of its efficacy in “in vitro” mechanical trial. The inner surface of the device is
mirrored and is based on physical concepts of light refraction and reflection. The
main advantage of such device is the reduced clinical time needed for bonding and
the lesser possibility of contamination during the process. 120 bodies of proof were
used for testing the shear bond strength of brackets bonded with the device. The
Adhesive Remnant Index (ARI) was also determined. The sample was divided into 2
groups. In group 1 a halogen light-curing unit was used and in group 2 a led lightcuring
unit was used. Each group was then subdivided. In subgroups H1 and L1, a
conventional light guide rod was used. In subgroups H2 and L2 bonding was
performed with the mirrored device coupled to the tip of the guide light rod. The
values obtained for the shear bond strength and the ARI in the subgroups were
compared. Results showed that there was no statistically significant difference for the
shear strength (p>0,05) and the ARI (p>0,05) between the subgroups. Conclusions
are that 1 - The tests of mechanical trials and the ARI analysis showed that the new
device fulfilled the requisites for bonding orthodontic accessories; 2 - time for bonding
was reduced to half, being necessary only one light exposure.

Identiferoai:union.ndltd.org:IBICT/oai:hermes.cpd.ufjf.br:ufjf/1579
Date14 February 2012
CreatorsMota Júnior, Sergio Luiz
ContributorsVitral, Robert Willer Farinazzo, Capelli Júnior, Jonas, Gravina, Marco Abdo
PublisherUniversidade Federal de Juiz de Fora, Programa de Pós-graduação em Saúde Brasileira, UFJF, Brasil, Faculdade de Medicina
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguagePortuguese
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/masterThesis
Sourcereponame:Repositório Institucional da UFJF, instname:Universidade Federal de Juiz de Fora, instacron:UFJF
Rightsinfo:eu-repo/semantics/openAccess

Page generated in 0.0033 seconds