Abstract
dc:description.abstractThe performance of an inductively coupled plasma mass spectrometer, ICP-MS, depends on the instrument's ability to transport sample ions through the vacuum interface and focus the ions into a well-defined beam that will eventually reach the mass analyzer. In this study two main experiments were performed on the Perkin Elmer NexION 300S, a commercial ICP-MS. First, planar laser-induced fluorescence images were taken of the ion beam in a working instrument downstream from a unique quadrupole ion deflector. The images showed the ability of the instrument design to focus the ions in the ion beam. Second, laser-induced fluorescence was used to characterize ion flow through the vacuum interface. The interface is unique to the NexION ICP-MS in that there are three extraction cones. The effect of a three-cone interface on ideal skimming is discussed.
Degree
thesis:*- Name thesis:degree_name
- MS
- Year dc:date
- 2017
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Larsen, Jessica Joline
Subjects
dc:subject × 4Rights
- Licence dc:rights.license
- Language dc:language
- english
Identifiers
dc:identifier.*- Repository record dc:identifier
- https://scholarsarchive.byu.edu/etd/6905