Ion Transport in a Commercial ICP-MS

Abstract

dc:description.abstract

The performance of an inductively coupled plasma mass spectrometer, ICP-MS, depends on the instrument's ability to transport sample ions through the vacuum interface and focus the ions into a well-defined beam that will eventually reach the mass analyzer. In this study two main experiments were performed on the Perkin Elmer NexION 300S, a commercial ICP-MS. First, planar laser-induced fluorescence images were taken of the ion beam in a working instrument downstream from a unique quadrupole ion deflector. The images showed the ability of the instrument design to focus the ions in the ion beam. Second, laser-induced fluorescence was used to characterize ion flow through the vacuum interface. The interface is unique to the NexION ICP-MS in that there are three extraction cones. The effect of a three-cone interface on ideal skimming is discussed.

Degree

thesis:*
Name thesis:degree_name
MS
Year dc:date
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Larsen, Jessica Joline

Subjects

dc:subject × 4

Rights

Licence dc:rights.license
Language dc:language
english

Identifiers

dc:identifier.*
Repository record dc:identifier
https://scholarsarchive.byu.edu/etd/6905

Chain of custody

source
Harvested from
Brigham Young University
Base URL
scholarsarchive.byu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation