Abstract
dc:description.abstractWith more young adults requiring total hip arthroplasties the need for bone saving implants becomes more important. The Zimmer Fitmore is a new bone saving implant that utilizes an implantation technique that reduces the damage to the muscle tissue allowing for patients to have a short recovery time as well as a new design that allows it to rest on the medial cortex. There has been anecdotal evidence that this device leads to early revision within six months of implantation due to failures occurring in the medial cortex. The main goal of this study was to computationally model the Zimmer Fitmore and compare it to the ML Taper to see if the failures are due to the design of the implant. The models were created using CT scans of the implants and the same implantation process was simulated for each. Two sizes for the cortical bone thickness, 4mm and 10mm, were used and contrasted with each other. The 10mm cortical thickness model showed that v the strains experienced by the Zimmer Fitmore femur were higher than that of the ML Taper. The 4mm model did not fully complete the simulation, but the results that were obtained showed an increased strain in Gruen zone 7. These results show that the design, not implantation method, could be to blame for the need for early revision when using the Zimmer Fitmore.
Degree
thesis:*- Name thesis:degree_name
- MS in Biomedical Engineering
- Discipline thesis:degree_discipline
- Biomedical and General Engineering
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Franklin, Tyler Kazuo
- Contributors dc:contributor
-
- Scott Hazelwood
Subjects
dc:subject × 11Identifiers
dc:identifier.*- Identifier
- 10.15368/theses.2013.29