Carleton University
Argon Calibration of a Cryogenic Distillation Column for Xenon Isotopic Enrichment
Abstract
dc:description.abstractThis work presents the calibration for a 14.6 m-high cryogenic distillation column which is performed using argon. This column is located at SNOLAB, an underground lab in Canada, and it is planned to study the potential of using distillation for xenon isotopic enrichment. Our study yields a measurement of the still’s height equivalent to theoretical plate (HETP) of 4.29±0.13 cm, consistent with a previous study using the same packing material. A new approach introduced in this analysis achieves a smaller error for this HETP compared to the previous study. Our HETP value will be used in the measurement of the vapour pressure isotopic effect of xenon in the future. In addition, we provide a measurement for the fractionation factor ln(α38−40) = (2.731± 0.017)×10−3 at the nominal boiling point, which has improved error and is consistent with theoretical values.
Degree
thesis:*- Name thesis:degree_name
- Master of Science (M.Sc.)
- Level thesis:degree_level
- Master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- Carleton University
- Year dc:date.issued
- 2023
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Emara, Abo-Bakr Mohamed Abdelfattah Mohamed
Subjects
dc:subject × 1Rights
dc:rights- Statement dc:rights
-
- Copyright © 2023 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, distribution to students, research and scholarship. Theses may only be shared by linking to the Carleton University Institutional Repository and no part may be copied without proper attribution to the author; no part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner.
- Language dc:language.iso