Contact Simulation for Evaluating Patient Specific Surgical Guide Stability

Abstract

dc:description.abstract

This study proposes a novel computational method to quantify guide stability for Patient Specific Instrumentation (PSI) guides. A finite element contact model was used to analyze the final position of PSI guides on a femur across a range of loading parameters representing forces applied by a surgeon during operative use. Separate segmentation methods were used for the guide and bone geometry to represent differences between segmentation and actual patient geometry. The region of loading parameters over which the guide exhibited a consistent final position was measured and reported as Guide Stability Score. The model was verified using cadaver specimens for which 3D printed PSI guides were applied to the knee using similar variations in guide loading. A strong correlation was found between the Guide Stability Score and the variance seen in the lab (R2 = 0.84), suggesting the model provides a useful tool for the evaluation of PSI guide stability.

Degree

thesis:*
Name thesis:degree_name
M.S.
Level thesis:degree_level
Masters Thesis

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Nierste, Vincent
Contributors dc:contributor
  • Chadd Clary
  • Paul Rullkoeter
  • Mei Yin

Subjects

dc:subject × 6

Rights

dc:rights
Statement dc:rights
  • <p>Copyright is held by the author. User is responsible for all copyright compliance.</p>
  • Vincent Nierste
Language dc:language
en

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.du.edu/etd/2147

Chain of custody

source
Harvested from
University of Denver
Base URL
digitalcommons.du.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Nierste, Vincent. Contact Simulation for Evaluating Patient Specific Surgical Guide Stability. Masters Thesis thesis, https://digitalcommons.du.edu/etd/2147