Abstract
dc:description.abstractThe blood draw portion of an in-home blood testing system was investigated with focus on ease of use, pain reduction, flow rate, and the total volume of blood drawn. Traditional style stainless steel hypodermic needles were used in order to investigate the effects of size, viscosity, surface tension, the presence of a porous medium, clogging by tissue, and surface coatings on flow rate through needles in the presence of simulated human capillary bed blood pressure. Non-coring point style needles were found to operate the best in terms of resistance to clogging. Surface effects were found in as large as 26 gauge (0.464 mm OD) needles at low pressure, and were prominent in 31 gauge (0.261 mm OD) and 33 gauge (0.210 mm OD) needles at high pressure. An array of 4-6 needles was conceived to enhance system reliability in light of complicating factors such as viscosity, random obstruction by porous medium, and surface tension.
Degree
thesis:*- Name thesis:degree_name
- M.S.
- Level thesis:degree_level
- Masters Thesis
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Huber, Nathan M.
- Contributors dc:contributor
-
- Corinne Lengsfeld
Subjects
dc:subject × 6Rights
dc:rights- Statement dc:rights
-
- <p>Copyright is held by the author. Permanently suppressed.</p>
- Nathan M. Huber
- Language dc:language
- en
Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.du.edu/etd/2188