East Tennessee State University
Development of a Portable Electroencephalogram with Display
Abstract
dc:description.abstractIn this experiment, the goal was to design, build, and test an Electroencephalogram (EEG) circuit that would then be paired with a touchscreen display as a portable unit for use in Schizophrenic research at East Tennessee State University at the Quillen College of Medicine. The requirements for this prototype denoted that the device had to be able to obtain a signal from two electrodes placed on the head of an individual with a third attached to the ankle as a ground, amplify and filter the signal, and display the results on a touchscreen display for recording and analyzing. The circuit design was approached using a basic three stage breakdown structure: Acquire, Amplify, Filter. This experiment outlines various lowpass and bandpass active filters used in conjunction with instrumentation amplifiers for signal acquisition and amplification. The final design is a 5th-order 25Hz Lowpass Filter EEG Circuit with variable Gain and Quality Factor.
Degree
thesis:*- Name thesis:degree_name
- MS (Master of Science)
- Level thesis:degree_level
- Thesis - unrestricted
- Discipline thesis:degree_discipline
- Engineering Technology
- Grantor dc:publisher
- East Tennessee State University
- Year dc:date.issued
- 2025
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Davis, Seth D
Subjects
dc:subject × 8Rights
dc:rights- Statement dc:rights
-
- Copyright by the authors.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Repository record dc:identifier
- https://dc.etsu.edu/etd/4540