Back to search

East Tennessee State University

Development of a Portable Electroencephalogram with Display

Abstract

dc:description.abstract

In this experiment, the goal was to design, build, and test an Electroencephalogram (EEG) circuit that would then be paired with a touchscreen display as a portable unit for use in Schizophrenic research at East Tennessee State University at the Quillen College of Medicine. The requirements for this prototype denoted that the device had to be able to obtain a signal from two electrodes placed on the head of an individual with a third attached to the ankle as a ground, amplify and filter the signal, and display the results on a touchscreen display for recording and analyzing. The circuit design was approached using a basic three stage breakdown structure: Acquire, Amplify, Filter. This experiment outlines various lowpass and bandpass active filters used in conjunction with instrumentation amplifiers for signal acquisition and amplification. The final design is a 5th-order 25Hz Lowpass Filter EEG Circuit with variable Gain and Quality Factor.

Degree

thesis:*
Name thesis:degree_name
MS (Master of Science)
Level thesis:degree_level
Thesis - unrestricted
Discipline thesis:degree_discipline
Engineering Technology
Grantor dc:publisher
East Tennessee State University
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Davis, Seth D

Subjects

dc:subject × 8

Rights

dc:rights
Statement dc:rights
  • Copyright by the authors.
Language dc:language
eng

Identifiers

dc:identifier.*
Repository record dc:identifier
https://dc.etsu.edu/etd/4540

Chain of custody

source
Harvested from
East Tennessee State University
Base URL
dc.etsu.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Davis, Seth D. Development of a Portable Electroencephalogram with Display. Thesis - unrestricted thesis, East Tennessee State University, 2025. https://dc.etsu.edu/etd/4540