Back to search

title="Search in OATD">Florida Atlantic University

MICROPROCESSOR BASED DRUG INFUSION CONTROL SYSTEM, EMPLOYING A MODEL REFERENCE ADAPTIVE CONTROL ALGORITHM, TO REGULATE BLOOD PRESSURE IN INTENSIVE CARE UNIT PATIENTS

Abstract

dc:description.abstract

This microprocessor based drug infusion control system is intended to aid doctors and nurses in the care of critically ill cardiac patients. The patient's arterial blood pressure is monitored and the infusion rate of the vasodilator sodium nitroprusside is regulated based on a model reference adaptive control algorithm. The algorithm employs a reference model to approximate the patient drug response. The reference model output is compared with the patient blood pressure change and the adaptive controller parameters are changed bringing the patient drug response in closer agreement with the reference model. Drug infusion is digitally controlled by a microprocessor based system and employs a stepper motor driven peristaltic pump. Simulation studies have validated the system. Animal experiments and clinical studies will be conducted later.

Degree

thesis:*
Level thesis:degree_level
Thesis

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • HERNANDEZ, LEO.

Subjects

dc:subject × 2

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.fau.edu/etd_general_3/561

Chain of custody

source
Harvested from
title="Search in OATD">Florida Atlantic University
Base URL
digitalcommons.fau.edu/do/oai/
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

HERNANDEZ, LEO.. MICROPROCESSOR BASED DRUG INFUSION CONTROL SYSTEM, EMPLOYING A MODEL REFERENCE ADAPTIVE CONTROL ALGORITHM, TO REGULATE BLOOD PRESSURE IN INTENSIVE CARE UNIT PATIENTS. Thesis thesis, https://digitalcommons.fau.edu/etd_general_3/561