Aristotle University Of Thessaloniki (AUTH)
Φθορά κινητού ένθετου πολυαιθυλενίου σε ολική αρθροπλαστική γόνατος: κλινική και πειραματική μελέτη
Abstract
dc:descriptionWe studied the wear of the Rotaglide mobile polyethylene bearings following experimental and clinical use. At 5 million cycles and with the same simulation system, the mobile bearings exhibited up to four times less wear than the fixed IB inserts, which were tested at the same time. We also studied 26 retrieved Rotaglide inserts with 8 years (3 months to 15.6 years) mean period of use. The 7mm and 9.5 mm of thickness inserts, showed moderate to severe wear after 9.5 years. In 51-85% of the upper surface, delamination dominates, but plastic deformation and pitting to a much lesser extend are also present. Little wear of burnishing type and abrasions, develop after 5.5 years mainly in the posterior condyles and in the thin inserts is accompanied by fracture and loss of material. In all bearings, regardless of thickness there is no significant backside wear. 18% of the backside surface had no wear, 42% showed some abrasions and scattered areas of burnishing and the remaining 40% is dominated by pitting inside areas of burnishing.
Degree
thesis:*- Grantor dc:publisher
- Aristotle University Of Thessaloniki (AUTH)
- Year dc:date
- 2008
Author and committee
dc:creator, dc:contributor.*- Authors dc:creator
-
- Nicolaides, A.
- Νικολαΐδης, Αναστάσιος
Subjects
dc:subject × 11Rights
- Language dc:language
- gre
Identifiers
dc:identifier.*- Identifier
- 10.12681/eadd/19384