University of Guelph
The Biomechanics of Locking Plates used in a Supercutaneous Fashion
Abstract
dc:description.abstractFracture fixation has progressed from perfect anatomic reduction to fracture stabilization preserving surrounding soft tissue. External fixators cause minimal soft tissue disruption during placement/removal. Despite these benefits, external fixators are bulky and costly leading to the investigation of alternatives. Locking plates, such as the string of pearls (SOP) and locking compression plate (LCP), gain strength from the screw-plate interface. Although soft tissue disruption is required for placement on the bone, precise contouring is not necessary. The present study compares the LCP and SOP placed 20mm off the bone surface to the external fixator. Constructs were tested in cyclic loading and compression to failure. The external fixator was statistically significantly stiffer (p=0.0050) than the LCP and SOP when placed supercutaneously. Additionally, the SOP was statistically stiffer than the LCP (p=0.0050). Thus, veterinary surgeons must carefully consider the implications of using locking in lieu of an external fixator.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Department thesis:degree_department
- Department of Clinical Studies
- Grantor dc:publisher
- University of Guelph
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Proctor, Ashley
- Advisor dc:contributor.advisor
-
- MacIver, Melissa
Subjects
dc:subject × 3Rights
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/10214/28620