NJIT
An intelligent computerized pulse-duplicator for the in-vitro evaluation of prosthetic heart valves
Abstract
dc:description.abstract<p>A computerized data acquisition system is presented for the in-vitro determination of the hydrodynamic performance of prosthetic heart valves. High speed data acquisition and ensemble averaging techniques are employed to minimize error and provide a standard for data collection and analysis across a wide range of simulated physiologic conditions. Processing algorithms automatically divide the cardiac cycle into phases based on landmarks in the pressure and flow waveforms, with minimal variability and without operator intervention. The contributions of valve resistance, regurgitation, ejection, leakage, and power loss to hydrodynamic performance are isolated for investigation. Phasic power loss analysis is reviewed as a relative measure of valve merit, with preliminary results revealing a need for further investigation into the clinical significance of such losses.</p>
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Biomedical Engineering - (M.S.)
- Discipline thesis:degree_discipline
- Biomedical Engineering Committee
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Andrews, John F.
- Contributors dc:contributor
-
- David S. Kristol
- Shlomo Gabbay
- Peter Engler
Subjects
dc:subject × 3Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.njit.edu/theses/3004