Monterey, CA; Naval Postgraduate School
A total absorption spectrometer for gamma rays
Abstract
dc:description.abstractA description of a total-absorption spectrometer for gamma rays is given. The spectrometer depends on the Cherenkov radiation emitted in 14 inches of lead glass by the shower electron-positron pairs formed by an incident gamma ray. The Cherenkov radiation is detected only when four photomultiplier s looking at the glass operate in coincidence. The added output of the photomultipliers is analyzed with a ten-channel pulse-height analyzer. The relative pulse-height output of the spectrometer varies linearly as the energy of incident monoenergetic electrons is varied from 42 Mev to 319 Mev. The energy resolution is about 30%. It is believed that the useful range of the spectrometer will extend to several Bev.
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Physics
- Level thesis:degree_level
- Masters
- Department dc:contributor.department
- Physics (PH)
- Grantor dc:publisher
- Monterey, CA; Naval Postgraduate School
- Year dc:date.issued
- 1955
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Jester, Malvern H. L.
Subjects
dc:subject × 1Identifiers
dc:identifier.*- OCLC
- ocm640316929