Design of a Precision, Adjustable Current Source for Use In an MRI-Based Robotic Catheter System

Abstract

dc:description

Heart catheter ablation surgeries often employ X-ray fluoroscopy to image the position of the catheter during the procedure. Magnetic Resonance Imaging (MRI) can be used to image the catheter position in real-time without exposing a patient to ionizing radiation while providing excellent soft tissue contrast. The MRI’s static magnetic field can be used in conjunction with current-carrying coils placed on a catheter to steer the catheter through a patient robotically. This thesis sees the development of a new system to precisely and accurately source current to these catheter coils in a way that ensures protection from the MRI’s harsh environment. The designed current driver boards with a switching topology allow for a significantly downsized system when compared to the older linear system. Relevant tests were performed on these boards to ensure their functionality and compliance with the MRI – including linearity, thermal drift, and filter performance.

Degree

thesis:*
Name thesis:degree_name
Master of Sciences (Engineering)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
EECS - Electrical Engineering
Grantor dc:publisher
Case Western Reserve University School of Graduate Studies
Year dc:date
2023

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Mousa, Cole M.
Contributors dc:contributor
  • Cavusoglu, M. Cenk

Subjects

dc:subject × 8

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Mousa, Cole M.. Design of a Precision, Adjustable Current Source for Use In an MRI-Based Robotic Catheter System. masters thesis, Case Western Reserve University School of Graduate Studies, 2023. http://rave.ohiolink.edu/etdc/view?acc_num=case1705085611465787