University of Toledo Health Science Campus
Development and implementation of quality-assurance standards for external beam intensity modulated radiation therapy
Abstract
dc:descriptionThe purpose of this project was to develop an effective and time-efficient plan checkingprocedure using a combination of ionization chambers and film. A 0.6 cc volume thimbleionization chamber, Kodak X-Omat V (XV) film and Kodak Extended Dose Range (EDR2) filmwere used. Three IMRT plans (one prostate, one head & neck and one brain) were validated.The plans were transferred from the patient geometry to a 30 cm x 30 cm x 20 cm solid waterphantom. The ionization chamber was calibrated and used to measure dose delivered to thephantom. The XV film was used to compare planar dose distributions and the EDR2 film wasused to compare axial dose distributions to the distributions computed by the treatment planningsystem. The tests set forth in this procedure provide adequate evaluation of IMRT treatmentplans.
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Biomedical Sciences (MSBS)
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- College of Medicine
- Grantor dc:publisher
- University of Toledo Health Science Campus
- Year dc:date
- 2009
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Hack, Joshua
- Contributors dc:contributor
-
- Parsai, E. Ishmael
Subjects
dc:subject × 3Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=mco1265034762