The Ohio State University
Design, engineering,and evaluation of a novel microgrid electrode array to monitor the electrical activity on the surface of the cerebral cortex
Abstract
dc:descriptionThe purpose of the study was to design and engineer a novel microdevice, a microgrid electrode array, that can further biomedical engineering research in the areas of neuroscience and neurology. Microfabrication techniques were developed to create the microdevice, which is capable of measuring electrical activity on the surface of the cerebral cortex at a greater level of spatial resolution than currently available for surface electrode grids. The device has been designed to detect the activity of individual cortical columns. Although penetrating microelectrodes are capable of detecting this level of detail, their invasiveness limits their use to human cortical tissue destined for surgical resection or strictly to animal studies. The development of a less-invasive non-penetrating microgrid electrode array suitable for human use would provide medical scientists with a unique and powerful tool for detailed brain mapping, gaining a better understanding of human cortical column function, as well as for exploring diseases and therapy options associated with the cerebral cortex. The fabrication protocol included several standard photolithography steps and many novel chemical and encapsulation methods. Characterization of the fabrication protocol steps and testing of the electrical connection integrity was performed using Scanning Electron Microscopy (SEM), optical microscopy, profilometery, and a microprobe station. Cyto-compatibility testing of the microgrid electrode with mammalian and human cortical cells consisted of qualitative and quantitative analyses. Evaluation with a porcine animal model consisted of collecting Visual Evoked Potentials (VEPs) from the primary visual cortex. Results from characterization of the fabrication process revealed that the desired geometries of the device were achieved. The resulting microgrid consisted of an array of surface contacts (200-micron platinum squares with bi-directional pitch of 400 microns) suspended in an insulating polydimethylsiloxane (PDMS) polymer. Cyto-compatibility testing with mammalian and human cells gave results that suggest that the device may be suitable for use with human patients. The results acquired from testing the device with an animal model revealed that the device was not only able to detect electrical activity from the cortex surface, but also suggests that activity corresponding to individual cortical columns can be acquired.
Degree
thesis:*- Name thesis:degree_name
- Doctor of Philosophy
- Level thesis:degree_level
- doctoral
- Discipline thesis:degree_discipline
- Biomedical Engineering
- Grantor dc:publisher
- The Ohio State University
- Year dc:date
- 2004
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Kitzmiller, Joseph Paul
- Advisor dc:contributor.advisor
-
- Hansford, Derek
Subjects
dc:subject × 4Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=osu1084824069