The Ohio State University
The micromotion between the polyethylene insert and the tibial tray in total knee prostheses
Abstract
dc:descriptionAseptic loosening is the primary means of failure of total knee replacements. Recent studies have suggested a causal relationship between polyethylene (PE) wear debris and aseptic loosening. A few retrieval studies of failed implants have indicated significant back surface wear, but attempts to measure this wear have been scarce. The primary emphasis of this study is the measurement of the relative motion between the insert and the tibial tray as a first step towards quantifying the wear at this interface. Linear Variable Differential Transducers (LVDTs) are used to measure the motion. The implants are placed under an axial cyclic load varying between 200 and 2500 Newtons for one million cycles. This loading is chosen to simulate the walking lad a 150 pound person experiences at the knee joint over a one year period. When collecting data points the frequency of the load is 0.5 Hertz, approximately walking speed...
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Biomedical Engineering
- Grantor dc:publisher
- The Ohio State University
- Year dc:date
- 1996
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Sosa, Miguel A., III
- Advisor dc:contributor.advisor
-
- Litsky, Alan
Subjects
dc:subject × 6Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=osu1100624700