The Ohio State University
Intraocular Pressure Characterization with Goldmann Applanation Tonometry and with a Take-Home Rebound Tonometer in Glaucoma Suspects
Abstract
dc:descriptionIntroduction:Glaucoma is a major cause of blindness throughout the world. Intraocular pressure (IOP) is the most important risk factor to aid in glaucoma diagnosis and to predict progression. IOP fluctuates throughout the day, usually peaking during nighttime; yet, most IOP measurements are made in eye clinics during the daytime using Goldmann applanation tonometry (GAT). Thus, peak IOP and fluctuations in IOP are not captured, forcing clinicians to analyze glaucoma risk with incomplete data. The subsequent inability to characterize a patient’s IOP variation is a critical barrier for the clinician to identify at-risk patients, who need early or aggressive IOP-based treatment. A take-home rebound tonometer (iCare HOME; iCare, Finland) can acquire IOPs over 24-hours, including non-business hours. The aim of this study is to test the hypothesis that a take-home rebound tonometer can provide a more complete assessment of the IOP profiles of glaucoma suspects than in-office GAT. Methods:Twenty-six participants (N = 26) were prospectively recruited into two cohorts. A case cohort (n = 13) comprised participants, who were glaucoma suspects due to suspicious optic nerves or who had ocular hypertension. A control cohort (n =13) comprised healthy participants. A retroactive chart review of all participants generated maximum IOP (Tmax) and minimum IOP (Tmin) for GAT measurements taken in- office. Each participant was required to have at least two previous in-office GAT measurements.All participants were then trained to perform self-tonometry using a take-home rebound tonometer (iCare HOME) and recorded their IOP five times a day spread across 24 hours for a 14-day period. Paired t-tests were used to compare Tmax, Tmin, and IOP range between the take-home rebound tonometer and GAT within groups, and non-paired t-tests to make inter-cohorts comparisons. Data from the left eye were arranged for cosinor analysis, which analyzed midline estimating statistic of rhythms (MESOR), amplitude, and acrophase outputs. Mean IOP and IOP standard deviation were also collected from the at-home rebound tonometer. The statistical significance threshold was set at α = 0.05.Results:Twenty-four subjects (mean age 49 ± 15 years for cases and 47 ± 16 years for controls) successfully performed enough measurements for statistical analysis. Inter-cohort comparisons for iCare HOME measurements found significantly higher MESOR, maximum IOP, minimum IOP, and mean IOP in the case cohort, compared to control cohort. No statistically significant differences were found between cohorts for amplitude, acrophase, IOP standard deviation, and IOP range. For GAT measurements, maximum IOP was significantly higher in the case cohort than in the control cohort, and no statistically significant differences between cohorts were present for minimum IOP and IOP range.Inter-device analysis between iCare HOME and GAT for the case cohort demonstrated significantly higher maximum IOP and IOP range when using the iCare HOME compared to GAT. Minimum IOP showed no statistically significant difference between devices. Linear regression analysis showed a significant correlation between the two devices for maximum IOP, minimum IOP, and IOP range. Bland-Altman analysis for the case cohort displayed a bias toward iCare HOME to measure greater than GAT for maximum IOP and IOP range with a slight bias toward GAT for minimum IOP.Inter-device analysis between iCare HOME and GAT for the control cohort demonstrated significantly higher maximum IOP and IOP range and significantly lower minimum IOP when measuring with iCare HOME compared to GAT. Linear regression analysis showed no significant correlations between iCare HOME and GAT measurements for maximum IOP, minimum IOP, and IOP range. Bland-Altman analysis showed similar results to case cohort with bias toward greater measurements with iCare HOME for maximum IOP and IOP range and a slight bias toward GAT for minimum IOP. Conclusion:The take-home rebound tonometer was able to detect a higher mean IOP and higher minimum IOP in the case cohort than in the control cohort. There were no differences in these parameters between the cohorts when IOP was measured with GAT during office hours, even though GAT was able to detect a larger IOP range in the case cohort than in the control cohort.Furthermore, take-home rebound tonometry measured a higher Tmax and a larger IOP range than GAT in both cohorts. Take-home rebound tonometry also detected a lower average Tmin than GAT in the case cohort. These results differ from several previous studies, which found that daytime IOP measured with GAT is higher than daytime IOP measured with rebound tonometry.All together our results suggest that at-home rebound tonometry, performed over 24 hours, may be better suited to detect high IOP in glaucoma suspects, compared to GAT measurements taken during office hours. This conclusion has direct clinical impact, as detecting high IOP is critical for assessing glaucoma risk and for setting a sufficient target IOP in patients who need treatment.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Vision Science
- Grantor dc:publisher
- The Ohio State University
- Year dc:date
- 2024
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Montgomery, Tyson Arthur
- Advisor dc:contributor.advisor
-
- Yuhas, Phillip
Subjects
dc:subject × 7Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=osu1732223971678005