University of Cincinnati
A BIOPARTICLE SEPARATION TECHNIQUE THROUGH MICROCHANNELS USING SEQUENTIAL PRESSURE PULSES
Abstract
dc:descriptionIn this work, an on-chip magnetic bead separator in an aqueous solution has been implemented on a plastic substrate. The technique has been successfully characterized in separating magnetic microspheres of 4.1 µm diameter from a suspension solution in DI water. µTAS (Micro Total Analysis System) frequently need to deal with bioparticle suspensions in solution. Separation of certain bioparticles is often desirable. Traditionally, methods like physical filtration have been used to separate biomolecules from suspension solutions. These methods, though well-established are not suited for integration on to mass-fabricated plastic lab-on-a-chip devices. The method developed in this work relies on the differential force exerted by application of high amplitude, short duration pressure pulses on a suspension solution, resulting in separation of suspended particles. From a dispensed volume of 500 nl of the suspension solution, up to 300 nl, or 60 % of volume has been cleared of particles.
Degree
thesis:*- Name thesis:degree_name
- MS
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Engineering : Electrical Engineering
- Grantor dc:publisher
- University of Cincinnati
- Year dc:date
- 2004
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- JAIN, ALOK
- Contributors dc:contributor
-
- Ahn, Dr. Chong
Subjects
dc:subject × 5Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=ucin1085759166