Abstract

dc:description

Microelectrodes implanted in the brain as transducers for neuronal impulses should ideally constitute dual functionality, i.e. stimulating and recording to avoid use of separate electrodes, smaller size for highly localized response, impedance stability for lesser power consumption, and non-metallic behavior to avoid MRI interaction. In this work a composite fiber microelectrode (140 µm), was developed, such that it has two conducting elements Carbon) well insulated and separated with Silicon Carbide (SiC), concentrically. In vitro studies were done to use the outer carbon layer as concentric ring stimulating electrode while recording from inner carbon core. An electrolytic etching technique was developed for fabricating electrode tips from SiC-C composite fiber. SEM was used to study carbon surface morphology, different tip geometries and shapes. Cyclic Voltammetry and Electrochemical Impedance Spectroscopy were used for the electrochemical characterization and modeling. The mechanics of redox and transport properties of the system were studied using ferrocene oxidation experiment. Effects of various stimulus waveforms and frequencies on charge transfer capability were studied. Different multielectrode configurations and designs were fabricated and tested for inductive charge development, electrical cross-talk and impedance change. Reversible charge injection limits were established using current and potential transient measurements, PH and temperature change, and direct observation of faradic processes. Biochemical sensing of neurotransmitters including Dopamine Hydrochloride and Vitamin C was done with the SiC-C composite electrodes and oxidation currents were found to vary linearly with concentration. In-Vivo action potential recordings from anesthetized rat’s brain with very high signal to noise ratio and firing response to external pressure stimulus were successfully obtained using novel SiC-C microelectrodes. The effectiveness of SiC-C composite microelectrodes, to electrically stimulate rat’s brain, was established by inserting these electrodes into motor cortex of rat’s brain and studying evoked motor functions.

Degree

thesis:*
Name thesis:degree_name
PhD
Level thesis:degree_level
doctoral
Discipline thesis:degree_discipline
Engineering : Materials Science
Grantor dc:publisher
University of Cincinnati
Year dc:date
2007

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Singh, Sherjang
Advisor dc:contributor.advisor
  • Buchanan, Dr. Relva

Subjects

dc:subject × 10

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Singh, Sherjang. SiC-C Composite Microelectrode for Biomedical Applications. doctoral thesis, University of Cincinnati, 2007. http://rave.ohiolink.edu/etdc/view?acc_num=ucin1170556512