Pohang University of Science and Technology
Microwave Power Module with Feedback System for Biomedical Plasma
Abstract
We designed a feedback system in a microwave plasma power module to reduce the variation of output power which can affect plasma temperature. In the feedback system, a power detector changed the coupled output power to DC voltage, and a voltage comparator compared the DC voltage to a reference voltage and then controlled the amplifier control voltage. To compare the difference between power module with feedback system and without feedback system, we checked the effect of the circuit temperature and reflected EMI on power module output and plasma temperature. We monitored the coupled power variation from the reference power (1 W) using a signal analyzer and measured the plasma temperature at 1-cm distance from the end of an argon plasma jet. With the feedback system, the power variation was 0.996 ~ 1.008 W/W and temperature variation was 40 ~ 40.5 °C for 5 min, whereas without the feedback system, power variation was 0.816 ~ 1.121 W/W and temperature variation was 38.9 ~ 42.4 °C. This feedback system will allow safe use of our plasma device in biomedical applications.
Author and committee
dc:creator, dc:contributor.*- Author
-
- 권순구
Identifiers
dc:identifier.*- Identifier
- hdl:2014.oak/2095
- Mods Identifier Other
- OAK-2014-01593