He3 cryostat for steady state nuclear magnetic resonance measurements in metals

Abstract

dc:description

A He³ cryostat was designed for steady state nuclear magnetic resonance measurements. The cryostat required about one liter of He³ at NTP. This amount of He³ was liquified at 1.2°K, which was the temperature of the surrounding liquid He⁴ bath, and then pumped on by means of a four-stage mercury diffusion pump. By this procedure, a liquid He³ temperature of 0.35 K was achieved and maintained for four hours. The r-f coil and sample were in direct contact with the liquid He³ to ensure sufficient heat transfer between the two. The temperature of the liquid He³ was measured by means of vapour pressure measurements of the evaporating He³ and by means of resistance measurements of a carbon resistance in contact with the liquid He³. The system was non-recirculating, since a total, uninterrupted run of four hours was considered long enough for most experiments. The He³ could also be retransferred within 20 minutes after such a run, and this process could be continued until in-sufficient quantities of liquid He⁴remained to cool the surroundings.

Degree

thesis:*
Name thesis:degree_name
Master of Applied Science - MASc
Level thesis:degree_level
master's
Discipline thesis:degree_discipline
Physics
Grantor dc:publisher
University of British Columbia
Year dc:date
1966

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Puls, Manfred Paul

Rights

dc:rights
Statement dc:rights
  • For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/2429/37817

Chain of custody

source
Harvested from
University of British Columbia
Base URL
circle.library.ubc.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Puls, Manfred Paul. He3 cryostat for steady state nuclear magnetic resonance measurements in metals. master's thesis, University of British Columbia, 1966. http://hdl.handle.net/2429/37817