Abstract
dc:descriptionA one dimensional buried channel CCD with a 30 /urn thick intrinsic layer was manufactured at TRIUMF for use as a radiation detector. In order to characterize the detector's response to radiation, six discrete X-Ray sources with energies ranging from 8 KeV to 60 keV were impinged on the device. An algorithm was developed to convert events detected by the CCD into spectra characteristic of each X-Ray source. The range of energies detectable and the energy resolution of the detector at different energies were studied.A model was developed and used to explain the shape of each spectrum. The spectrum from an 241Am source was also studied to determine the detector's response to a multi-energetic source.
Degree
thesis:*- Name thesis:degree_name
- Master of Science - MSc
- Level thesis:degree_level
- master's
- Discipline thesis:degree_discipline
- Physics
- Grantor dc:publisher
- University of British Columbia
- Year dc:date
- 1997
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Patten, Scott
Rights
dc:rights- Statement dc:rights
-
- For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
- Language dc:language
- eng
Identifiers
dc:identifier.*- Handle dc:identifier
- http://hdl.handle.net/2429/6576